Owon Hds2102s Firmware Update Fix May 2026

For Users:

For Manufacturers (suggested product improvements):

Updating the Owon HDS2102S Go to product viewer dialog for this item.

firmware is a critical fix for a known bug where the 2ns/div horizontal setting incorrectly displays waveforms (effectively acting as 2.5ns/div). 🛠️ Update Process

Follow these steps to apply the fix using the Mass Storage Class (MSC) method: Check Hardware Compatibility

Navigate to System > F2 (System) > Next Page to view your current version.

Warning: Firmware is serial-number specific. Do not attempt to flash if your device uses a read-only Gigadevice flash chip (often identified by "NOT upgradable" status in community archives). Download Firmware

Visit the Owon Download Center to find the correct .upp file for your model and serial range. Prepare the Device Connect the scope to your PC via USB. Go to System > Next Page and set USB to MSC.

The device will appear as a removable disk on your computer. Flash the Fix Copy the Scope.upp file to the root of the removable disk. Power the scope OFF and then ON.

The device will automatically upgrade and shut down when finished. ⚠️ Important Fixes & Risks

2ns/div Accuracy: Updating to v1.5.1 or later specifically resolves the graticule spacing error for high-frequency signals.

Inversion Bug: Some users report the screen appearing upside down after certain updates; ensure you match your exact hardware revision to avoid this.

Stability: If the update hangs, try using a shorter USB cable to ensure a stable data connection. If you'd like, let me know: Your current firmware version (e.g., V1.x, V4.x) The first four digits of your serial number If you are experiencing the screen orientation issue

The OWON HDS2102S is a popular 3-in-1 handheld oscilloscope, but early versions often shipped with bugs that hampered its 100MHz potential. Specifically, users frequently encounter a graticule display issue at the 2ns/div setting and missing RMS measurements in the multimeter or oscilloscope menus. Fortunately, a firmware update can resolve these bugs and improve stability. Key Issues Fixed by Firmware Updates

Updating your HDS2102S to the latest version (such as V1.5.1 or later) addresses several critical performance gaps:

2ns/div Horizontal Scaling: Early firmware had a display discrepancy where the waveform did not align correctly with the graticules at its fastest sweep speed. Version 1.5.1 fixes this, ensuring a 50 MHz signal correctly occupies 10 divisions.

RMS Measurement Support: Some hardware versions lacked standard RMS measurements in the "Measure" menu. Newer firmware (typically 4.7.1+) restores or renames these commands to provide accurate readings.

USB Connectivity Stability: Updates often improve the handshake between the device and PC software, ensuring the "Mass Storage Class" (MSC) mode works reliably for file transfers. Step-by-Step Firmware Update Guide

Before starting, ensure your device is fully charged or connected to power. A power failure during a flash can brick the unit. 1. Download the Correct Firmware

Visit the official OWON Download Center and select the HDS200 Series.

Verify Hardware Version: It is critical to match the firmware to your hardware. Check your current version in the System > Info menu on your scope. owon hds2102s firmware update fix

Download the Package: Look for the specific HDS2102S update file (often a .upp or .bin file). 2. Prepare the Device for Transfer

The HDS2102S uses a simple "drag-and-drop" method for updates by appearing as a removable drive on your PC. Turn on the oscilloscope. Press the System button.

Select F2 (System), navigate to the next page, and select Upgrade. Connect the scope to your PC via the USB-C cable.

In the System menu, ensure the USB mode is set to MSC (Mass Storage Class). 3. Perform the Flash

When the scope appears as a removable disk on your PC, copy the downloaded firmware file (e.g., Scope.upp) directly to the root directory.

Once the transfer is complete, safely eject the drive from your PC.

Power Cycle: Turn the scope off and then back on. The device will detect the file, begin the upgrade automatically, and then shut down once finished.

Turn the scope back on and verify the new version number in the System menu. Troubleshooting Common Update Failures

Upside-Down Screen: If your display appears flipped after an update, you likely flashed firmware meant for a different hardware revision. You must find the specific firmware version compatible with your hardware serial number to revert the change.

Device Not Recognized: If your PC doesn't see the scope, try a different USB-C cable (some are "charge only") or ensure you have manually switched the device to MSC mode in the menu.

"Update Failed" Message: Ensure the firmware filename matches exactly what the device expects (usually Scope.upp). Do not put the file inside a folder on the device. Hardware Version Recommended Firmware Key Change V1.0 Fixes 2ns/div graticule bug V3.0 General stability and UI fixes V4.0+ Includes RMS measurements

For more advanced support or archived firmware files, the EEVblog Forums maintain a community-driven repository of various hardware-specific versions.

The Owon HDS2102S had a well-documented "glitch" where the display showed incorrect timing at the high-speed 2ns/division setting. Specifically, a 50 MHz signal (which should have a 20ns period) was incorrectly rendered as having a 16ns period.

Here is the "story" of how this was fixed and the current status of updates: The Problem & The Fix

The Glitch: Users noticed the horizontal graticules were incorrectly spaced at the fastest timebase, leading to inaccurate waveform measurements.

The Patch: Owon released Firmware Version 1.5.1 specifically to address this display discrepancy. After applying this update, the 50 MHz signal correctly fills the expected 10 divisions (at 2ns/div), restoring accuracy to the handheld unit. Important: Check Your Version First

Before attempting an update, you must identify your current firmware series. Owon uses different hardware revisions that are not cross-compatible:

V1.x.x or V2.x.x: These are earlier versions. Version 1.5.1 is the known stable fix for the display bug in this series.

V3.x.x or V4.x.x: These are newer hardware revisions. Owon has explicitly stated that users with these versions should not attempt to "upgrade" to V1.x.x or V2.x.x, as it can brick the device. Currently, there are often no downloadable updates for these series because the bug was fixed at the factory level. How to Update (General Steps) If you have a V1 or V2 unit and need the fix:

Contact Support: Because the official website often removes older firmware files, it is best to contact Owon Support directly to request the specific .bin file for your serial number. For Users:

Preparation: Use a small-capacity USB drive (formatted to FAT32). The Process:

Place the firmware file in the root directory of the USB drive. Connect it to the HDS2102S using the USB-C adapter. Access the System Menu and select the Update option.

Do not power off the device during the process, as this will permanently damage the internal software.

To verify your current version, go to Menu > System > About on your device.

The OWON HDS2102S handheld oscilloscope (part of the HDS200 series) famously had a display scaling bug where the timebase actually measured as . This caused high-frequency signals, such as a wave, to appear with incorrect periods (showing instead of the correct The Fix: Firmware Version 1.5.1

OWON released firmware version 1.5.1 to resolve this specific horizontal spacing issue. Users who have updated to this version or later report that the

range now correctly displays waveforms with accurate timing and graticule spacing. Update Procedure Overview

While OWON does not always provide a public "one-click" download for all serial numbers, the general process for updating an HDS200 series device involves:

Identify Current Version: Check your current firmware in the device settings menu to see if you are below v1.5.1.

Acquire Firmware: Visit the OWON Download Center to look for the latest software and firmware packages specific to your model.

Note: Some users obtain firmware files from community archives like GitHub repositories or by contacting OWON Support directly if their serial number is not listed.

Flash via USB: Most HDS200 series updates are performed by connecting the device to a PC via USB and using a specific "DSO Flash" or "Upgrade" utility provided in the firmware zip folder.

Verify: After the update, confirm the version is 1.5.1 or higher and test a known-frequency signal at the Key Improvements in v1.5.1 Timebase Accuracy: Corrected the display discrepancy.

Graticule Alignment: Ensured signal periods align correctly with the on-screen grid.

Stability: General bug fixes reported by the community on forums like EEVblog.

Here are a few options for a post about the Owon HDS2102S firmware update fix, tailored to different platforms.

If your screen is completely black and the key combos do nothing, you have a bootloader corruption. You need serial recovery via the micro-USB port using a TTL-to-USB adapter (3.3V logic, e.g., FT232RL).

If you have attempted the SD card rescue (Part 2) and the USB DFU rescue (Part 3), yet the device still only shows a dim backlight, the SPI flash memory chip (Winbond W25Q64) may have physically failed. OWON uses a consumer-grade flash chip that can die after about 10,000 write cycles. One corrupted update can fry the first sector.

In this case, the "fix" is a hardware rework: Desolder the old flash chip, program a new one with a universal programmer (TL866II Plus), and re-solder it. This is a $50 fix at a repair shop versus buying a new $350 scope.

Final Verdict: The OWON HDS2102S is a capable device let down by fragile update infrastructure. By following the forced recovery steps outlined above, you can fix 90% of "bricked" units. Remember: Format FAT32, use the F1 key trick, and never, ever unplug during write. the screen remains completely black)

Have you successfully fixed your HDS2102S? Share your OS version and the specific error message in the comments below to help the next engineer facing the "owon hds2102s firmware update fix" nightmare.

This assumes the device is stuck, fails to boot, or shows erratic behavior after an update attempt, and provides a structured fix procedure.


Recovery success correlated strongly with: availability of correct firmware image, ability to trigger bootloader, and presence of an intact bootloader.

(Community forums and vendor documentation were consulted; specific URLs and forum threads are omitted per instructions.)

Acknowledgments Thanks to community contributors who shared recovery notes and to lab technicians who assisted with hardware procedures.

The Owon HDS2102S had a notorious "glitch" where signals at the 2ns/div setting appeared stretched or incorrectly scaled, effectively lying to the user about their frequency. This was finally resolved with the release of Firmware v1.51. The Core Fix: Accuracy at 2ns/div

Before the fix, a 50 MHz signal displayed at the fastest horizontal scale (2ns/div) showed one full cycle taking up eight divisions—an incorrect 16ns period.

The Update: After flashing v1.51, the same signal correctly shows half a cycle taking up five divisions, confirming the accurate 10ns duration for a 50 MHz waveform.

Why it mattered: Without this fix, the device's highest-speed captures were mathematically unreliable for high-frequency measurements. How to Apply the Update

The update process for the HDS200 series uses a "Mass Storage Class" (MSC) method rather than a dedicated flash tool:

Preparation: Download the correct firmware file (typically named Scope.upp) from the Lilliput (OWON) Download Center.

Enable Upgrade Mode: On the scope, press System, navigate to the second page of the System menu (usually F2), and select Upgrade.

USB Connection: Connect the scope to your PC. Change the USB setting to MSC. A removable disk drive will appear on your computer.

Transfer & Reboot: Copy the Scope.upp file directly onto the removable disk. Power the scope off and then back on.

Completion: The scope will automatically run the upgrade and shut down when finished. Troubleshooting & Version Confusion

"Unupgradable" Units: Some early hardware versions (identifiable by serial numbers) used Gigadevice flash chips with OTP SecureRom, which may prevent firmware updates entirely.

The Version Gap: Users have reported receiving new units with versions like v4.3.0. If your device has a version higher than v1.5.1, you likely already have the fix and should not "downgrade" to older v1.x firmware files.

DMM Performance: Note that switching to DMM or Function Generator mode can sometimes slow down data acquisition speeds; a full restart usually restores performance. To see the before-and-after comparison of the display fix:

The Owon HDS2102S firmware update, specifically version 1.5.1 or higher, resolves a significant display scaling error in the 2ns/div horizontal range. Later versions, such as v4.7.1, introduce Ampere units for clamp meter functionality, though updates are hardware-locked and require matching the correct firmware to the specific unit. For firmware downloads, visit the Owon Download Center


  • Copy the .upg file to the root directory of the USB drive.
  • Rename file if needed – some OWON models require the exact name:
    HDS2102S.upg (check your device’s update instruction).
    When in doubt, use the original downloaded name first.
  • If the SD card method above does not work (i.e., the screen remains completely black), your bootloader is corrupted. You need a hardware fix using USB and a jumper.

    Warning: This requires opening the casing, which voids your warranty.